iCare Tonovet Plus

Icare® TONOVET Plus is the new generation tonometer for quick and easy veterinary IOP measuring.

The Icare® TONOVET Plus tonometer is used for measuring IOP on animal patients by general veterinary practitioners, veterinary ophthalmologists and other veterinary medical personnel.

Brands:ICARE

Description

As with all Icare® tonometers, the TONOVET Plus tonometer is based on a rebound measuring principle, in which a light weight probe is used to make momentary contact with the cornea. The gentle measurement does not require any anesthetic nor does it cause corneal reflex in the majority of animals. Icare® rebound tonometers are easy to use as they do not require maintenance calibration by the user.

The new Icare TONOVET Plus is even easier to use.  It features an advanced user

Features

 

  • Bigger OLED display in colours
  • Informative user interface
    • Visual illustration on measuring sequence
    • Clear error messages
  • One press measurement option
    • Also possible as single measurements
  • Deviation indication in colours
    • Green = no/very low standard deviation
    • Yellow = slight standard deviation
    • Red = high standard deviation
  • New, updated calibrations for 4 species = 4 settings:
    • Dog – Canine
    • Cat – Feline w Separated from the dog setting
    • Rabbit – Lupine w ALL NEW
    • Horse – Equine
  • New calibrations have been developed in co-operation with the University of Wisconsin-Madison, the Iowa State University and the University of Georgia.
  • Option to adjust the volume and turn the lights off.
    • When positioning lights are turned off, they’ll appear on the display.
  • Intelligent Positioning Assistant Lights

 

Model name TV011
Dimensions 24 – 29 mm (W) x 35 – 95 mm (H) x 215 mm (L)
Weight 140 g (without batteries), 230 g (4 x AA batteries)
Power supply 4 x AA non-rechargeable batteries, 1.5V alkaline LR6
Measurement range 10 – 60 mmHg
Accuracy ± 2.5 mmHg (10 – 30 mmHg) and ± 10% (> 30 – 60 mmHg)
Repeatability (coefficient of variation) < 8 %
Accuracy of display 1 mmHg
Display unit Millimeter of mercury (mmHg)
This device has BF-type electric shock protection

Additional information

Brand

ICARE

Model

TV011